Publications
Download:
File size:
1927 kb
Format:
application/pdf
Author:
Sandén, Bengt (Uppsala University, Department of Surgical Sciences)
Title:
Fixation of Spinal Implants: Clinical and Experimental Studies on the Effects of Hydroxyapatite Coating
Department:
Uppsala University, Medicinska vetenskapsområdet, Faculty of Medicine, Department of Surgical Sciences
Publication type:
Doctoral thesis, comprehensive summary (Other academic)
Language:
English
Place of publ.:
Uppsala
Publisher:
Acta Universitatis Upsaliensis
Pages:
48
Series:
Comprehensive Summaries of Uppsala Dissertations from the Faculty of Medicine, ISSN 0282-7476; 1085
Year of publ.:
2001
URI:
urn:nbn:se:uu:diva-1422
Permanent link:
http://urn.kb.se/resolve?urn=urn:nbn:se:uu:diva-1422
ISBN:
91-554-5137-3
Subject category:
Surgery
Research subject:
Orthopaedics
Keywords(en) :
Surgery, Hydroxyapatite, pedicle screw, histomorphometry, fixation, radiolucent zones, screw loosening
Keywords(sv) :
Kirurgi
Abstract(en) :

Loosening of implants is a frequent complication in orthopaedic surgery. The aim of this thesis was to evaluate hydroxyapatite coating of pedicle screws with mechanical and histological methods and to correlate the results with the radiological findings.

Hydroxyapatite (HA) coating of pedicle screws was investigated in experimental studies in sheep. Nine sheep were operated on with destabilising laminectomies at L2-L3 and L4-L5 and stabilised with pedicle screw instrumentations, where HA coated and uncoated screws were used in a randomised fashion. After both 6 and 12 weeks of implantation, the HA coated screws demonstrated a higher bone-to-implant contact and a higher amount of bone in the area close to the screw. The pull-out resistance, stiffness and total energy to failure was higher for the HA coated screws after 12 weeks of implantation.

HA coating of pedicle screws was examined in two clinical studies. In the first series, the patients were stabilised with pedicle screw instrumentations, where HA coated screws were used in either the upper or lower instrumented level in a randomised fashion. After 10-22 months of implantation, the extraction torque was markedly higher for the HA coated screws. In the second series, instrumentations with uncoated, partly HA coated or fully HA coated screws were used. After 11-16 months implantation, the extraction torques were recorded. There were significant differences between all three groups, with the lowest extraction torques for the uncoated screws and the highest torques for the fully HA coated screws. The frequency of radiolucent zones surrounding the screws was higher for the uncoated screws than for the HA coated screws.

Radiographs from both experimental and clinical studies were examined. Screws demonstrating radiolucent zones were compared to screws without zones with respect to pull-out resistance, extraction torque, bone-to-implant contact and amount of bone surrounding the screws. All these variables demonstrated higher values for the screws without radiolucent zones. The frequency of radiolucent zones surrounding uncoated screws in the clinical study was 53%.

Conclusions: Radiolucent zones are good predictors of screw loosening. The frequency of radiolucent zones is higher than previously described. Hydroxyapatite coating improves the purchase of pedicle screws and reduces the frequency of screw loosening.

Public defence:
2001-10-26, Rosénsalen, Akademiska sjukhuset, ingång 95/96, Uppsala, 13:15
Degree:
degree of Doctor of Philosophy (Faculty of Medicine)