Description and performance of an electron-ion coincidence TOF spectrometer used at the Brazilian synchrotron facility LNLS
Show others and affiliations
2010 (English)In: Journal of Electron Spectroscopy and Related Phenomena, ISSN 0368-2048, Vol. 180, no 1-3, 6-13 p.Article in journal (Refereed) Published
Abstract [en]

This paper reports the characteristics and performance of a Time-of-Flight Mass Spectrometer (TOF-MS) for coincidence measurements between electrons and ions that has been developed jointly in Sweden and Brazil. The spectrometer, used for studies of inner-shell photoexcitation of molecules in the gas-phase, has been optimized by implementing ion and electron lenses to allow the use of relatively small diameter detectors. Simulations were performed to understand the lens performance and they show that ions (electrons) could be collected without angular discrimination with a maximum kinetic energy up to ten (two) times higher than without the lens actions. A rotary vacuum chamber allows the spectrometer axis to be positioned at different angles relative to the polarization vector of the excitation beam. An important characteristic of the apparatus is that the acquisition setup allows a multi-hit capability with 1 ns resolution. Hereby, Photoelectron-Photoion-Photoion Coincidence (PEPIPICO) measurements can be performed on molecules containing two or more atoms of equal mass. A method to obtain experimental detection efficiencies of a single ion and one of one, two or three electrons has been developed. A systematic study of the interaction region has been performed to determine the shape of the photon and gas beams. Measurements on molecular nitrogen demonstrate the spectrometer's ability to resolve fragments with the same charge to mass ratio arriving within only a few ns. Simulations and experimental results of fragmentation of two singly charged cation nitrogen atoms agree, confirming that the spectrometer performance is well understood. (C) 2010 Elsevier B.V. All rights reserved.

Place, publisher, year, edition, pages
2010. Vol. 180, no 1-3, 6-13 p.
Keyword [en]
Multicoincidence, Instrumentation, Time-of-Flight, Synchrotron
National Category
Physical Sciences
URN: urn:nbn:se:uu:diva-135815DOI: 10.1016/j.elspec.2010.02.007ISI: 000279612800002OAI: oai:DiVA.org:uu-135815DiVA: diva2:375832
Available from: 2010-12-09 Created: 2010-12-08 Last updated: 2011-02-23Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full text
By organisation
Surface and Interface Science
In the same journal
Journal of Electron Spectroscopy and Related Phenomena
Physical Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 265 hits
ReferencesLink to record
Permanent link

Direct link