Evaluation of Gentian cystatin C reagent on Abbott Ci8200 and calculation of glomerular filtration rate expressed in mL/min/1.73 m(2) from the cystatin C values in mg/L
Uppsala University, Disciplinary Domain of Medicine and Pharmacy, Faculty of Medicine, Department of Medical Sciences. (Biokemisk struktur och funktion)
(Biokemisk struktur och funktion)
Uppsala University, Disciplinary Domain of Medicine and Pharmacy, Faculty of Medicine, Department of Medical Sciences. (Biokemisk struktur och funktion)
(Biokemisk struktur och funktion)
Show others and affiliations
2007 (English)In: Scandinavian Journal of Clinical and Laboratory Investigation, ISSN 0036-5513, E-ISSN 1502-7686, Vol. 67, no 5, 560-567 p.Article in journal (Refereed) Published
Abstract [en]

OBJECTIVE: Estimation of the glomerular filtration rate (GFR) is essential when evaluating patients with kidney disease and treating patients with drugs eliminated from the circulation by the kidneys. Cystatin C has been shown in several studies to be superior to creatinine in the estimation of GFR. At our hospitals, there is an increasing demand for cystatin C and at present we perform approximately 1500 cystatin C analyses a month. We thus need the assay available 24 h/day and to have it on our routine chemistry instrument to minimize handling time per test and time to reported test results. MATERIAL AND METHODS: We have evaluated a new cystatin C immunoassay from Gentian (Gentian, Moss, Norway) on Architect ci8200 (Abbott Laboratories, Abbott Park, Ill., USA). A prerequisite at our hospital is that cystatin C results are reported as a calculated GFR in mL/min/1.73 m(2), so we also made a comparison with iohexol clearance. RESULTS: The Gentian cystatin C assay showed good agreement with the corresponding assay from Dade Behring (Deerfield, Ill., USA) and good inter-laboratory concordance. The assay has very low total imprecision, good linearity and strong correlation with iohexol clearance (R (2) = 0.956). The equation for the correlation curve is: y = 79.901x(-1.4389). CONCLUSIONS: There was low inter-laboratory variation between the three laboratories involved in the cystatin C evaluation, and thus all three laboratories can use the same equation for calculating the estimated GFR.

Place, publisher, year, edition, pages
2007. Vol. 67, no 5, 560-567 p.
Keyword [en]
Cystatin C, glomerular filtration ratio, human, iohexol, kidney
National Category
Medical and Health Sciences
Identifiers
URN: urn:nbn:se:uu:diva-12976DOI: 10.1080/00365510601187773ISI: 000249067400013PubMedID: 17763193OAI: oai:DiVA.org:uu-12976DiVA: diva2:40746
Available from: 2008-01-19 Created: 2008-01-19 Last updated: 2017-12-11Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Authority records BETA

Larsson, Anders

Search in DiVA

By author/editor
Larsson, Anders
By organisation
Department of Medical SciencesClinical Chemistry
In the same journal
Scandinavian Journal of Clinical and Laboratory Investigation
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 429 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf