Multicenter clinical assessment of the raumedic Neurovent-P intracranial pressure sensor: A report by the brainIT group
Show others and affiliations
2008 (English)In: Neurosurgery, ISSN 0148-396X, E-ISSN 1524-4040, Vol. 63, no 6, 1152-1158 p.Article in journal (Refereed) Published
Abstract [en]

OBJECTIVE: The aim of this study was to evaluate the robustness and zero-drift of an intracranial pressure sensor, Neurovent-P (Raumedic AG, Munchberg, Germany), when used in the clinical environment. METHODS: A prospective multicenter trial, conforming to the International Organization for Standardization 14155 Standard, was conducted in 6 European BrainIT centers between July 2005 and December 2006. Ninety-nine catheters were used. The study was observational, followed by a centralized sensor bench test after catheter removal. RESULTS: The mean recorded value before probe insertion was 0.17 +/- 1.1 mm Hg. Readings outside the range 1 mm Hg were recorded in only 3 centers on a total of 15 catheters. Complications were minimal and mainly related to the insertion bolt. The mean recorded pressure value at removal was 0.8 +/- 2.2 mm Hg. No relationship was identified between postremoval reading and length of monitoring. The postremoval bench test indicated the probability of a system failure, defined as a drift of more than 3 mm Hg, at a range between 12 and 17%. CONCLUSION: The Neurovent-P catheter performed well in clinical use in terms of robustness. The majority of technical complications were associated with the bolt fixation technology. Adverse events were rare and clinically nonsignificant. Despite the earlier reported excellent bench test zero-drift rates, under the more demanding clinical conditions, zero-drift rate remains a concern with catheter tip strain gauge technology. This performance is similar, and not superior, to other intracranial pressure devices.

Place, publisher, year, edition, pages
2008. Vol. 63, no 6, 1152-1158 p.
Keyword [en]
Intracranial pressure, Multicenter trials, Physiological monitoring, Technology assessment
National Category
Medical and Health Sciences
URN: urn:nbn:se:uu:diva-152153DOI: 10.1227/01.NEU.0000335148.87042.D7ISI: 000261643400045PubMedID: 19057328OAI: oai:DiVA.org:uu-152153DiVA: diva2:412692
Available from: 2011-04-26 Created: 2011-04-26 Last updated: 2011-04-26Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed
By organisation
In the same journal
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 152 hits
ReferencesLink to record
Permanent link

Direct link