Comparison of Nasal Continuous Positive Airway Pressure Delivered by Seven Ventilators Using Simulated Neonatal Breathing
Uppsala University, Disciplinary Domain of Medicine and Pharmacy, Faculty of Medicine, Department of Surgical Sciences, Anaesthesiology and Intensive Care.
2013 (English)In: Pediatric Critical Care Medicine, ISSN 1529-7535, E-ISSN 1947-3893, Vol. 14, no 4, E196-E201 p.Article in journal (Refereed) Published
Abstract [en]

Objectives: Nasal continuous positive airway pressure (NCPAP) is an established treatment for respiratory distress in neonates. Most modern ventilators are able to provide NCPAP. There have been no large studies examining the properties of NCPAP delivered by ventilators. The aim of this study was to compare pressure stability and imposed work of breathing (iWOB) for NCPAP delivered by ventilators using simulated neonatal breathing. Design: Experimental in vitro study. Setting: Research laboratory in Sweden. Intervention: None. Measurements and Main Results: Neonatal breathing was simulated using a mechanical lung simulator. Seven ventilators were tested at different CPAP levels using two breath profiles. Pressure stability and iWOB were determined. Results from three ventilators revealed that they provided a slight pressure support. For these ventilators, iWOB could not be calculated. There were large differences in pressure stability and iWOB between the tested ventilators. For simulations using the 3.4-kg breath profile, the pressure swings around the mean pressure were more than five times greater, and iWOB more than four times higher, for the system with the highest measured values compared with the system with the lowest. Overall, the Fabian ventilator was the most pressure stable system. Evita XL and SERVO-i were found more pressure stable than Fabian in some simulations. The results for iWOB were in accordance with pressure stability for systems that allowed determination of this variable. Conclusions: Some of the tested ventilators unexpectedly provided a minor degree of pressure support. In terms of pressure stability, we have not found any advantages of ventilators as a group compared with Bubble CPAP, Neopuff, and variable flow generators that were tested in our previous study. The variation between individual systems is great within both categories. The clinical importance of these findings needs further investigation.

Place, publisher, year, edition, pages
2013. Vol. 14, no 4, E196-E201 p.
Keyword [en]
continuous positive airway pressure, equipment design, infant, newborn, positive-pressure, respiration/instrumentation, work of breathing
National Category
Medical and Health Sciences
URN: urn:nbn:se:uu:diva-202453DOI: 10.1097/PCC.0b013e31827212e4ISI: 000318680000005OAI: oai:DiVA.org:uu-202453DiVA: diva2:634975
Available from: 2013-07-02 Created: 2013-06-24 Last updated: 2016-02-29Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full text

Search in DiVA

By author/editor
Zetterström, Henrik
By organisation
Anaesthesiology and Intensive Care
In the same journal
Pediatric Critical Care Medicine
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 263 hits
ReferencesLink to record
Permanent link

Direct link