uu.seUppsala University Publications
Change search
ReferencesLink to record
Permanent link

Direct link
A Monte Carlo study of the dependence of early frame sampling on uncertainty and bias in pharmacokinetic parameters from dynamic PET.
Show others and affiliations
2015 (English)In: Journal of Nuclear Medicine Technology, ISSN 0091-4916, E-ISSN 1535-5675, Vol. 43, no 1Article in journal (Refereed) Published
Abstract [en]

UNLABELLED: Compartmental modeling of dynamic PET data enables quantification of tracer kinetics in vivo, through the calculated model parameters. In this study, we aimed to investigate the effect of early frame sampling and reconstruction method on pharmacokinetic parameters obtained from a 2-tissue model, in terms of bias and uncertainty (SD).

METHODS: The GATE Monte Carlo software was used to simulate 2 × 15 dynamic 3'-deoxy-3'-(18)F-fluorothymidine ((18)F-FLT) brain PET studies, typical in terms of noise level and kinetic parameters. The data were reconstructed by both 3-dimensional (3D) filtered backprojection with reprojection (3DRP) and 3D ordered-subset expectation maximization (OSEM) into 6 dynamic image sets with different early frame durations of 1, 2, 4, 6, 10, and 15 s. Bias and SD were evaluated for fitted parameter estimates, calculated from regions of interest.

RESULTS: The 2-tissue-model parameter estimates K1, k2, and fraction of arterial blood in tissue depended on early frame sampling, and a sampling of 6-15 s generally minimized bias and SD. The shortest sampling of 1 s yielded a 25% and 42% larger bias than the other schemes, for 3DRP and OSEM, respectively, and a parameter uncertainty that was 10%-70% higher. The schemes from 4 to 15 s were generally not significantly different in regards to bias and SD. Typically, the reconstruction method 3DRP yielded less frame-sampling dependence and less uncertain results, compared with OSEM, but was on average more biased.

CONCLUSION: Of the 6 sampling schemes investigated in this study, an early frame duration of 6-15 s generally kept both bias and uncertainty to a minimum, for both 3DRP and OSEM reconstructions. Very-short frames of 1 s should be avoided because they typically resulted in the largest parameter bias and uncertainty. Furthermore, 3DRP may be preferred over OSEM for short frames with poor statistics.

Place, publisher, year, edition, pages
2015. Vol. 43, no 1
National Category
Medical and Health Sciences
URN: urn:nbn:se:uu:diva-286933DOI: 10.2967/jnmt.114.141754PubMedID: 25613339OAI: oai:DiVA.org:uu-286933DiVA: diva2:922191
Available from: 2016-04-22 Created: 2016-04-22 Last updated: 2016-04-22

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Sörensen, Jens
By organisation
In the same journal
Journal of Nuclear Medicine Technology
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar

Altmetric score

Total: 88 hits
ReferencesLink to record
Permanent link

Direct link