The Roche Immunoturbidimetric Albumin Method on Cobas c 501 Gives Higher Values Than the Abbott and Roche BCP Methods When Analyzing Patient Plasma Samples.
Uppsala University, Disciplinary Domain of Medicine and Pharmacy, Faculty of Medicine, Department of Medical Sciences, Biochemial structure and function.
Uppsala University, Disciplinary Domain of Medicine and Pharmacy, Faculty of Medicine, Department of Medical Sciences, Biochemial structure and function.
Uppsala University, Disciplinary Domain of Medicine and Pharmacy, Faculty of Medicine, Department of Medical Sciences, Biochemial structure and function.
Show others and affiliations
2016 (English)In: Journal of clinical laboratory analysis (Print), ISSN 0887-8013, E-ISSN 1098-2825Article in journal (Refereed) Epub ahead of print
Abstract [en]

BACKGROUND: Serum/plasma albumin is an important and widely used laboratory marker and it is important that we measure albumin correctly without bias. We had indications that the immunoturbidimetric method on Cobas c 501 and the bromocresol purple (BCP) method on Architect 16000 differed, so we decided to study these methods more closely.

METHOD: A total of 1,951 patient requests with albumin measured with both the Architect BCP and Cobas immunoturbidimetric methods were extracted from the laboratory system. A comparison with fresh plasma samples was also performed that included immunoturbidimetric and BCP methods on Cobas c 501 and analysis of the international protein calibrator ERM-DA470k/IFCC.

RESULTS: The median difference between the Abbott BCP and Roche immunoturbidimetric methods was 3.3 g/l and the Roche method overestimated ERM-DA470k/IFCC by 2.2 g/l. The Roche immunoturbidimetric method gave higher values than the Roche BCP method: y = 1.111x - 0.739, R² = 0.971.

CONCLUSION: The Roche immunoturbidimetric albumin method gives clearly higher values than the Abbott and Roche BCP methods when analyzing fresh patient samples. The differences between the two methods were similar at normal and low albumin levels.

Place, publisher, year, edition, pages
2016.
National Category
Clinical Laboratory Medicine
Identifiers
URN: urn:nbn:se:uu:diva-296007DOI: 10.1002/jcla.21921PubMedID: 27169354OAI: oai:DiVA.org:uu-296007DiVA: diva2:935800
Available from: 2016-06-12 Created: 2016-06-12 Last updated: 2016-06-12

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed
By organisation
Biochemial structure and function
In the same journal
Journal of clinical laboratory analysis (Print)
Clinical Laboratory Medicine

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 38 hits
ReferencesLink to record
Permanent link

Direct link