Analytical and clinical evaluation of the Bayer ADVIA Centaur automated B-type natriuretic peptide assay in patients with heart failure: a multisite study
Show others and affiliations
2004 (English)In: Clinical Chemistry, ISSN 0009-9147, E-ISSN 1530-8561, Vol. 50, no 5, 867-873 p.Article in journal (Refereed) Published
Abstract [en]


B-Type natriuretic peptide (BNP) is released from the left ventricle of the heart into the circulation in response to ventricular stretching and volume overload. Increased BNP concentrations are associated with heart failure (HF).


We evaluated the analytical and clinical performance of the Bayer ADVIA Centaur BNP assay. Studies included precision, analytical correlation (against the Shionogi ShionoRIA and Biosite Triage BNP assays), BNP results for blood collected in plastic tubes containing EDTA vs other collection tubes, high-dose hook effect, detection limits, and interferences. The clinical performance was tested on 2243 blood samples collected from 983 apparently healthy individuals, 538 patients with chronic disease but without HF (renal insufficiency, chronic obstructive pulmonary disease, diabetes, and hypertension), and 722 patients with HF (New York Heart Association classes I-IV).


The ADVIA Centaur assay had total imprecision (CV) of 3.4%, 2.9%, and 2.4% at BNP concentrations of 48, 461, and 1768 ng/L, respectively. The Passing-Bablok correlations to the ShionoRIA and Triage were as follows: ADVIA Centaur = 1.11(ShionoRIA) - 1.19 ng/L (r = 0.98); ADVIA Centaur = 0.78(Triage) + 5.89 ng/L (r = 0.92), respectively. Of the different blood collection tubes, only EDTA plastic tubes (with and without the barrier gel) were acceptable. The lower detection limit was 0.5 ng/L, and there were no interferences from common analytes, other neuropeptides, or unusual antibodies. BNP exhibited different reference intervals according to age and gender. BNP concentrations increased progressively as the severity of HF increased.


The ADVIA Centaur is the first commercially available BNP assay for use on an automated immunochemistry platform. This assay has good analytical and clinical performance characteristics for diagnosing HF.

Place, publisher, year, edition, pages
2004. Vol. 50, no 5, 867-873 p.
Keyword [en]
Adult, Aged, Aged; 80 and over, Autoanalysis, Female, Heart Failure; Congestive/*diagnosis, Humans, Male, Middle Aged, Natriuretic Peptide; Brain/*blood, ROC Curve, Reference Values, Sensitivity and Specificity
National Category
Medical and Health Sciences
URN: urn:nbn:se:uu:diva-71574DOI: 10.1373/clinchem.2003.026138PubMedID: 15010423OAI: oai:DiVA.org:uu-71574DiVA: diva2:99485
Available from: 2005-09-13 Created: 2005-09-13 Last updated: 2012-07-24Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMedhttp://www.ncbi.nlm.nih.gov/entrez/query.fcgi?db=PubMed&cmd=Retrieve&list_uids=15010423&dopt=Citation

Search in DiVA

By author/editor
Wallentin, LarsJohnston, Nina
By organisation
UCR-Uppsala Clinical Research CenterCardiology
In the same journal
Clinical Chemistry
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 124 hits
ReferencesLink to record
Permanent link

Direct link