Data-Driven Methods for Microwave Sensor Devices in Musculoskeletal Diagnostics
Uppsala University, Disciplinary Domain of Science and Technology, Technology, Department of Electrical Engineering, Solid-State Electronics.ORCID iD: 0000-0001-8065-0094
2024 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Microwave sensors can be used within medicine as they use non-ionizing radiation, are often low cost, and can be designed for a specific purpose. The application of microwave sensors for diagnostics and monitoring can be improved using appropriate data analysis. The multi-layered structure of the human body makes the measurements on people complex. A tremendous effort is required to create an analytical model of the body. In this context a data-driven approach, building a model that learns from previous measurements, is more suitable to analyze the data. This thesis aims to address statistical and data-driven approaches based on microwave sensor data for biomedical applications.

A significant part of this thesis deals with microwave sensors for assessing muscle quality. It details the progress from initial clinical campaign to the creation of a machine learning algorithm to assess the local body composition. Such a device would be suitable for screening age-related muscle disorders like sarcopenia and muscle atrophy. Statistical analysis following the initial clinical campaign revealed no significant differences in the microwave data. Therefore, new sensor designs were evaluated. The most promising sensor was used in a small clinical campaign where it was able to detect a change in muscle size for one patient with multiple measurements over time. Successive measurements followed on tissue emulating phantoms and volunteers. For data analysis a machine learning algorithm was designed to predict the skin, fat, and muscle properties. This changes the aim from assessing muscle quality to assessing local body composition. For phantom data the algorithm was accurate for skin and fat and for volunteer data for fat and muscle. Crucially, the algorithm also performed better with more data available, meaning that results should improve if more data is collected.

Microwave sensors have also been employed to assess bone. The first of two applications was to monitor the bone healing progression post surgery treating craniosynostosis. No substantial conclusions could be drawn from the statistical analysis most likely due to measurement uncertainties. The second application used a purpose-built setup for controlled measurements in ex vivo bone samples submerged in liquid, to simulate an in vivo environment. The purpose was to estimate the dielectric properties of bone. The derived bone properties were lower than expected, probably due to air trapped inside the sample.

Place, publisher, year, edition, pages
Uppsala: Acta Universitatis Upsaliensis, 2024. , p. 93
Series
Digital Comprehensive Summaries of Uppsala Dissertations from the Faculty of Science and Technology, ISSN 1651-6214 ; 2359
Keywords [en]
Machine Learning, Microwave Sensors, Data-Driven Modeling, Statistical Analysis
Keywords [sv]
Maskininlärning, Mikrovågssensorer, Datadriven modellering, Statistisk Analys
National Category
Signal Processing Other Computer and Information Science Other Medical Engineering
Research subject
Engineering Science with specialization in Electronics
Identifiers
URN: urn:nbn:se:uu:diva-521537ISBN: 978-91-513-2021-2 (print)OAI: oai:DiVA.org:uu-521537DiVA, id: diva2:1831304
Public defence
2024-03-15, lecture room Heinz-Otto Kreiss, Lägerhyddsvägen 1, Uppsala, 09:00 (English)
Opponent
Supervisors
Available from: 2024-02-20 Created: 2024-01-25 Last updated: 2026-06-01
List of papers
1. MAS: Standalone Microwave Resonator to Assess Muscle Quality
Open this publication in new window or tab >>MAS: Standalone Microwave Resonator to Assess Muscle Quality
Show others...
2021 (English)In: Sensors, E-ISSN 1424-8220, Vol. 21, no 16, article id 5485Article in journal (Refereed) Published
Abstract [en]

Microwave-based sensing for tissue analysis is recently gaining interest due to advantages such as non-ionizing radiation and non-invasiveness. We have developed a set of transmission sensors for microwave-based real-time sensing to quantify muscle mass and quality. In connection, we verified the sensors by 3D simulations, tested them in a laboratory on a homogeneous three-layer tissue model, and collected pilot clinical data in 20 patients and 25 healthy volunteers. This report focuses on initial sensor designs for the Muscle Analyzer System (MAS), their simulation, laboratory trials and clinical trials followed by developing three new sensors and their performance comparison. In the clinical studies, correlation studies were done to compare MAS performance with other clinical standards, specifically the skeletal muscle index, for muscle mass quantification. The results showed limited signal penetration depth for the Split Ring Resonator (SRR) sensor. New sensors were designed incorporating Substrate Integrated Waveguides (SIW) and a bandstop filter to overcome this problem. The sensors were validated through 3D simulations in which they showed increased penetration depth through tissue when compared to the SRR. The second-generation sensors offer higher penetration depth which will improve clinical data collection and validation. The bandstop filter is fabricated and studied in a group of volunteers, showing more reliable data that warrants further continuation of this development.

Place, publisher, year, edition, pages
MDPIMDPI AG, 2021
Keywords
muscle quality, sarcopenia, microwave sensing, bandstop filter, substrate integrated waveguides
National Category
Medical Laboratory Technologies
Identifiers
urn:nbn:se:uu:diva-455509 (URN)10.3390/s21165485 (DOI)000690024700001 ()34450927 (PubMedID)
Available from: 2021-10-07 Created: 2021-10-07 Last updated: 2025-02-09Bibliographically approved
2. Muscle Analyzer System: Exploring Correlation Between Novel Microwave Resonator and Ultrasound-based Tissue Information in the Thigh
Open this publication in new window or tab >>Muscle Analyzer System: Exploring Correlation Between Novel Microwave Resonator and Ultrasound-based Tissue Information in the Thigh
Show others...
2022 (English)In: 2022 16th European Conference on Antennas and Propagation (EuCAP), Institute of Electrical and Electronics Engineers (IEEE), 2022, article id 9769216Conference paper, Published paper (Refereed)
Abstract [en]

A microwave sensor to safely measure quality of muscle tissue for diagnosis and screening of diseases and medical conditions characterized by fat infiltration in muscle is presented. Fat infiltration in muscle may be seen by a lower dielectric constant of muscle at microwave frequencies corresponding to the large contrast between fat and muscle tissues. A planar resonator based on a bandstop filter and optimized to noninvasively interrogate muscle in the thigh on tissue quality is proposed. Currently, a study based on clinical trials is carried out, and, here, we present a preliminary correlation between skin and fat thicknesses and rectus femoris cross sectional area (CSA) measured with ultrasound and the proposed sensor's resonance frequency. CST simulations based on the ultrasound information guide the analysis. We see that although there are signs of a potential correlation between CSA and resonance, skin and fat variability is still an issue to overcome.

Place, publisher, year, edition, pages
Institute of Electrical and Electronics Engineers (IEEE), 2022
Keywords
Sensors, Muscle Quality, Bandstop Filter, Clinical Measurements
National Category
Other Electrical Engineering, Electronic Engineering, Information Engineering Medical Instrumentation
Identifiers
urn:nbn:se:uu:diva-466134 (URN)10.23919/EuCAP53622.2022.9769216 (DOI)000815113901017 ()978-88-31299-04-6 (ISBN)978-1-6654-1604-7 (ISBN)
Conference
16th European Conference on Antennas and Propagation (EuCAP) 2022, 27 March-1 April, Madrid, Spain
Available from: 2022-01-24 Created: 2022-01-24 Last updated: 2026-09-18Bibliographically approved
3. Machine Learning Algorithm to Extract Properties of ATE Phantoms from Microwave Measurements
Open this publication in new window or tab >>Machine Learning Algorithm to Extract Properties of ATE Phantoms from Microwave Measurements
2024 (English)In: International journal of microwave and wireless technologies, ISSN 1759-0795, E-ISSN 1759-0787, Vol. 16, no 10, p. 1624-1631Article in journal (Refereed) Published
Abstract [en]

The Muscle Analyzer System (MAS) project wants to create a standalone microwave device that can assess the muscle quality, called the MAS device. To achieve that an algorithm that can derive the properties of skin, fat and muscle from the measurements is needed. This paper presents a machine learning algorithm that aims to do precisely that. The algorithm relies on first predicting the skin using the data from the MAS device, then predicting the fat again using the data from the MAS but also the predicted skin value and lastly the muscle is predicted using the microwave data together with the skin and fat predictions. Data have been collected in phantom experiments, materials that mimick the dielectric properties of human tissues. The algorithm is trained to predict the properties of said phantoms. The results show that the prediction for skin thickness works well, the fat thickness prediction is okay but the muscle prediction struggles. This is partly due to the error from the skin and fat layers are propagated to the muscle layer and partly because the muscle layer is farthest away from the sensor, which makes getting information from that layer harder.

Place, publisher, year, edition, pages
Cambridge University Press, 2024
National Category
Signal Processing Medical Instrumentation
Identifiers
urn:nbn:se:uu:diva-521534 (URN)10.1017/S1759078724000102 (DOI)001157395400001 ()2-s2.0-85184581625 (Scopus ID)
Available from: 2024-01-25 Created: 2024-01-25 Last updated: 2026-02-26Bibliographically approved
4. Machine Learning Powered Microwave Device for Local Body Composition Assessment
Open this publication in new window or tab >>Machine Learning Powered Microwave Device for Local Body Composition Assessment
Show others...
2024 (English)In: IEEE Sensors Journal, ISSN 1530-437X, E-ISSN 1558-1748, Vol. 24, no 5, p. 7030-7041Article in journal (Refereed) Published
Abstract [en]

In this paper a standalone microwave device is evaluated for its ability to assess local body composition with the ultimate goal to assess muscle quality. Data have been collected from volunteers who were measured on their thigh using the microwave device and ultrasound. A machine learning algorithm with three stages is designed that utilizes the stacked nature of the tissues in the thigh to predict skin and fat thickness and the cross-sectional area of the rectus femoris muscle. The input to the algorithm is the signal response from the microwave sensor and also the prediction from the previous layers. The ultrasound measurements are used as the ground truth labels for each tissue to train the machine learning models. The measurements were performed with two sensors, where usage of the combined data from both sensors produced the best results for fat and muscle, 0.57 and 0.63 in R 2 score, respectively. In the drop analysis, a step where a select proportion of the data is temporarily removed, the identified models showed increased scores with a larger amount of data available indicating the learning of the models improves with more data. Although the results are encouraging more data is ultimately needed to further study the algorithm.

Place, publisher, year, edition, pages
Institute of Electrical and Electronics Engineers (IEEE), 2024
Keywords
Muscles;Sensors;Fats;Phantoms;Ultrasonic variables measurement;Skin;Machine learning;Biomedical signal processing;Machine learning;Microwave sensor;Signal analysis
National Category
Signal Processing Medical Instrumentation
Identifiers
urn:nbn:se:uu:diva-521533 (URN)10.1109/JSEN.2023.3344581 (DOI)001280078400001 ()2-s2.0-85181565370 (Scopus ID)
Available from: 2024-01-25 Created: 2024-01-25 Last updated: 2025-02-20Bibliographically approved
5. Evaluation of Microwave Sensor for Bone Healing Monitoring after Craniosynostosis Surgery
Open this publication in new window or tab >>Evaluation of Microwave Sensor for Bone Healing Monitoring after Craniosynostosis Surgery
Show others...
(English)Manuscript (preprint) (Other academic)
National Category
Signal Processing Other Medical Engineering
Identifiers
urn:nbn:se:uu:diva-521535 (URN)
Available from: 2024-01-25 Created: 2024-01-25 Last updated: 2026-04-23
6. Setup to Perform Ex Vivo Coaxial Transmission Bone Measurements in a Simulated In Vivo Environment
Open this publication in new window or tab >>Setup to Perform Ex Vivo Coaxial Transmission Bone Measurements in a Simulated In Vivo Environment
(English)Manuscript (preprint) (Other academic)
National Category
Other Electrical Engineering, Electronic Engineering, Information Engineering Other Medical Engineering
Identifiers
urn:nbn:se:uu:diva-521536 (URN)
Available from: 2024-01-25 Created: 2024-01-25 Last updated: 2024-01-25

Open Access in DiVA

UUThesis_Mattsson,V-2024(2900 kB)908 downloads
File information
File name FULLTEXT01.pdfFile size 2900 kBChecksum SHA-512
fd3a3e4338eb5b1d6bdd0b6ccd2e3733dbd0dfafc969399849343b2f8cb3f7553de053da16d46bd2005f8a20810d3ab7c862f3ae1afe1bae6a4ec30b42e3bb2f
Type fulltextMimetype application/pdf

Authority records

Mattsson, Viktor

Search in DiVA

By author/editor
Mattsson, Viktor
By organisation
Solid-State Electronics
Signal ProcessingOther Computer and Information ScienceOther Medical Engineering

Search outside of DiVA

GoogleGoogle Scholar
Total: 911 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

isbn
urn-nbn

Altmetric score

isbn
urn-nbn
Total: 1959 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf