NOVIFAST: A Fast Algorithm for Accurate and Precise VFA MRIT1MappingShow others and affiliations
2018 (English)In: IEEE Transactions on Medical Imaging, ISSN 0278-0062, E-ISSN 1558-254X, Vol. 37, no 11, p. 2414-2427Article in journal (Refereed) Published
Abstract [en]
In quantitative magnetic resonance T 1 mapping, the variable flip angle (VFA) steady state spoiled gradient recalled echo (SPGR) imaging technique is popular as it provides a series of high resolution T 1 weighted images in a clinically feasible time. Fast, linear methods that estimate T 1 maps from these weighted images have been proposed, such as DESPOT1 and iterative re-weighted linear least squares. More accurate, non-linear least squares (NLLS) estimators are in play, but these are generally much slower and require careful initialization. In this paper, we present NOVIFAST, a novel NLLS-based algorithm specifically tailored to VFA SPGR T 1 mapping. By exploiting the particular structure of the SPGR model, a computationally efficient, yet accurate and precise T 1 map estimator is derived. Simulation and in vivo human brain experiments demonstrate a twenty-fold speed gain of NOVIFAST compared with conventional gradient-based NLLS estimators while maintaining a high precision and accuracy. Moreover, NOVIFAST is eight times faster than the efficient implementations of the variable projection (VARPRO) method. Furthermore, NOVIFAST is shown to be robust against initialization.
Place, publisher, year, edition, pages
IEEE, 2018. Vol. 37, no 11, p. 2414-2427
Keywords [en]
T-1 mapping, variable flip angle, SPGR, DESPOT1, relaxometry
National Category
Signal Processing
Identifiers
URN: urn:nbn:se:uu:diva-366258DOI: 10.1109/TMI.2018.2833288ISI: 000449113800004PubMedID: 29993537OAI: oai:DiVA.org:uu-366258DiVA, id: diva2:1264050
Funder
EU, FP7, Seventh Framework Programme, FP7/2007-2013The European Space Agency (ESA)2018-11-192018-11-192019-01-07Bibliographically approved