Automated GMP-production of α-[11 C]methyl-L-tryptophan using a tracer production system (TPS)Show others and affiliations
2018 (English)In: Journal of labelled compounds & radiopharmaceuticals, ISSN 0362-4803, E-ISSN 1099-1344, Vol. 61, no 14, p. 1106-1109Article in journal (Refereed) Published
Abstract [en]
The radiosynthesis and GMP validation of [11 C]AMT for human use are described. Three consecutive batches were produced giving 940-3790 MBq (4%-17% RCY, decay corrected, based on [11 C]CO2 ) of the tracer. The molar activity at the end of synthesis was 19 to 35 GBq/μmol, the radiochemical purity was ≥98%, and the enantiomeric purity was >99%. While the synthesis method was automated using a new generation of synthesis equipment, tracer production system developed in house, the method should be readily applicable to other synthesis platforms with minor modifications.
Place, publisher, year, edition, pages
2018. Vol. 61, no 14, p. 1106-1109
Keywords [en]
[11C]AMT, automation, methyl iodide, positron emission tomography, tracer production system
National Category
Medicinal Chemistry
Identifiers
URN: urn:nbn:se:uu:diva-368809DOI: 10.1002/jlcr.3648ISI: 000453709100011PubMedID: 29902836OAI: oai:DiVA.org:uu-368809DiVA, id: diva2:1268997
2018-12-072018-12-072019-01-15Bibliographically approved